32
STEINDORFF
®
PROBLEM
REASON
SOLUTION
IV.
Viewing tube:
The two eyes’ field of view is
different
The interpupillar distance is not
correct
Adjust the interpupillar distance
The diopter is not right
Adjust the diopter
Not adapt to the microscope
observation
When look into the objective, do
not stare at the specimen but at the
whole field of view, or move the
eyes away to see other things, then
back into