STEINDORFF
®
2.2.9.
Connecting the power cord (Figure 12, 13 and 14)
Do not bring the power cord to bear a powerful stress.
When being bent or wrapping, the cable and wires will
be broken easily.
1.
Turn the main switch
①
on “O” (off) state before
connecting the power cord.
2.
Insert the plugs
②
into the power jack
③
of the
microscope safely.
3.
Plug the power cord
④
into the power supply receptacle.
Make sure the connection is well.
4.
Insert aviatic BNC connector plugs
⑨
into aviatic BNC
connector jack
⑧
.
Do use the supplied power cord all the time. If lost or
damaged, select the same standard cord.
Connect the power cord correctly, to ensure the
instrument is earthing.
2.2.10.
Replacing the fuse (Figure 12and 13)
Do remember to turn the main switch
①
on the state of “O” (off)
before replacing the fuse, and unplug the power cord. Rotate the
fuse
⑥
kits out of the holder
⑦
by the “--” type screwdriver,
replace a new fuse, then rotate back to the holder again.
Fuse rating: 250V, 500mA.