Hardness (∑
alkaline earth
ions)
≤0.02 mmol/l
≤0.02 mmol/l
Appearance
Colourless, clear, without impurities
CAUTION: using distilled water which does not meet the above
parameters will cause malfunction of the unit and may even lead to
its damage!
Use only deionized water with a maximum conductivity of 15 μs/cm
-
higher values may lead to unit failure.
There are special "ultra pure" waters available on the market
dedicated for sterilization applications - the use of such waters is
recommended.
3.3.5
Emptying the waste water tank
If during the operation of the unit the message "FULL" appears on the
display next to the WATER1 position, this means that the waste water
tank is full and should be emptied. To do so, connect a hose to the
waste water drainage connector (15). The waste water will be drained
when it reaches a maximum temperature of 70
o
C - if it is higher than
this check that the fan is working correctly.
CAUTION: hot water - risk of scalding.
3.3.6
Placing items on the trays
Open the door (3) and using the tray holder (19) remove the
respective tray to place the items to be sterilized on it.
CAUTION: only pull out/place trays using the supplied tray handle as
they may be hot -
risk of burns!
IMPORTANT:
•
when sterilizing items of different materials at the same time, they
should be grouped on a given tray according to the same material
i.e. do not place different materials on the same tray.
•
There should be free space between items in the same tray - they
must not touch each other.