Turning the unit on Then press the parameter setting menu
button (a) on the control panel to access the following mode
setting submenu:
•
"UNWRAPPED..." - disposable vacuum sterilization of unwrapped
items; selectable at 120
o
C or 134
o
C; operating pressure 110 KPa or
210 KPa; duration of sterilization cycle at 121
o
C - 20 minutes, while
at 134
o
C - 4 minutes; drying time at 121
o
C - 5 minutes, while at
134
o
C
-
12 minutes.
•
"WRAPPED..." - triple vacuum sterilization of packaged
items; choice of temperature 121
o
C or 134
o
C; working
pressure 110 KPa or 210 KPa; duration of sterilization cycle
at 121
o
C - 20 minutes and at 134
o
C - 4 minutes; drying time
at 121
o
C - 5 minutes, at 134
o
C - 12 minutes.
•
"WRAPPED (B&D) - control sterilization according to the Bowie
and Dick method. This is an automatic program testing the
efficiency of the unit, conducted at 134
o
C.
NOTE: for this program a special test package according to Bowie
and Dick method is needed. Such a package is not included in the
unit!
•
"COTTON" - triple sterilization of cotton materials; working
temperature 121
o
C; working pressure 110 KPa; duration of the
cycles 15 minutes; drying time 15 minutes.