4 – 20
Remote Operation
•
OFSL
SLOPE6DB
•
OMOD
LOCKIN
•
OFSE
OFF
•
OFST
0.0
•
KCLK
ON
•
ALRM
ON
4.4.13
Status commands
Lock status
LOCK?
Reads the current value of the reference oscillator lock status. Returns
a token, with values
z
=
(
UNLOCKED 0
,
LOCKED 1
, or
NOTPLL 2
).
If the SR124 reference oscillator is operating in one of the external
reference modes (
FMOD
EXT1F
,
FMOD
EXT2F
, or
FMOD
EXT3F
), then
the query
LOCK?
will return either
UNLOCKED 0
or
LOCKED 1
based
on the current status.
If the SR124 reference oscillator is configured for either
FMOD
INTERNAL
or
FMOD
RVCO
, then
LOCK?
responds with
NOTPLL
2
.
LOCK?
Example:
LOCKED
Overload
OVLD?
Reads the current value of the signal overload status. Returns an
integer between 1 and 15 if an overload is detected, or 0 if there is no
overload.
The response integer is binary-weighted based on the four (4) sepa-
rate signal stages that can generate an overload detect:
•
1 Preamp overload
•
2 Current amp overload
•
4 Intermediate (AC) amp overload
•
8 Output (DC) amp overload
For example, if the preamp and the intermediate amp are both over-
loading,
OVLD?
will respond with
5
(
=
1
+
4).
OVLD?
Example:
0
SR124
Analog Lock-In Amplifier