4 – 14
Remote Operation
Output filter time constant
OFLT(?)
{
z
}
Set (query) the output filter time constant setting
{
to
z
}
. Allowable
values are:
z
Value
Full-scale sensitivity
TCMIN 0
<
500
µ
s
TC1MS 1
1 ms
TC3MS 2
3 ms
TC10MS 3
10 ms
TC30MS 4
30 ms
TC100MS 5
100 ms
TC300MS 6
300 ms
TC1S 7
1 s
TC3S 8
3 s
TC10S 9
10 s
TC30S 10
30 s
TC100S 11
100 s
TC300S 12
300 s
OFLT 7
Example:
Output filter slope
OFSL(?)
{
z
}
Set (query) the output filter slope (rollo
ff
)
{
to
z
=
(
SLOPE6DB 0
,
SLOPE12DB 1
)
}
.
OFSL SLOPE12DB
Example:
4.4.8
Output commands
Output mode
OMOD(?)
{
z
}
Set (query) the output mode
{
to
z
=
(
LOCKIN 0
,
ACVOLT 1
)
}
.
OMOD?
Example:
LOCKIN
Output o
ff
set enable
OFSE(?)
{
z
}
Set (query) the output o
ff
set mode
{
to
z
=
(
OFF 0
,
ON 1
)
}
.
OFSE 1
Example:
SR124