4.4
Commands
4 – 15
Output o
ff
set magnitude
OFST(?)
{
g
}
Set (query) the output o
ff
set magnitude
{
to
g
}
, in percent full scale.
The default value is
OFST
0.00
.
The allowed range for
g
is
−
1000
≤
g
≤
+
1000 (
±
10
×
full-scale).
OFST 300.253
Example:
4.4.9
Setup commands
Key clicks
KCLK(?)
{
z
}
Set (query) audible key clicks
{
to
z
=
(
OFF 0
,
ON 1
)
}
.
Note there is no corresponding front-panel method to access this
command; it is exclusive to the remote interface.
Audible alarms
ALRM(?)
{
z
}
Set (query) audible alarms
{
to
z
=
(
OFF 0
,
ON 1
)
}
.
Note that all sounds that are not “key clicks” are considered “alarms”
for the purpose of the
ALRM
command. There is no corresponding
front-panel methdo to access this command; it is exclusive to the
remote interface.
Save user settings
SSET(?)
{
z
}
Save (query) the user settings
{
to non-volatile block
z
=
(
USER0 0
,
USER1 1
, . . . ,
USER8 8
)
}
.
The set version of
SSET
saves user settings to parameter block
z
.
The query version
SSET?
returns the most block name or number
z
most recently saved into.
Recall user settings
RSET(?)
{
z
}
Retrieve (query) the user settings
{
to non-volatile block
z
=
(
USER0 0
,
USER1 1
, . . . ,
USER8 8
,
DEFAULT 9
)
}
.
The set version of
RSET
retreives user settings from non-volatile
block
z
and reconfigures the SR124 to those settings. The query
form simply returns the name
/
number of the most-recently retreived
memory block.
Note that
RSET
DEFAULT
is equivalent to
*RST
.
SR124
Analog Lock-In Amplifier