General Information
Safety and Preparation for Use
Dangerous voltages, capable of causing injury or death, are
WARNING
present in this instrument. Do not remove the product covers
or panels. Do not apply power or operate the product without
all covers and panels in place.
AC line voltage
The SR124 Analog Lock-In Amplifier operates from a 100 V, 120 V,
220 V, or 240 V nominal AC power source having a line frequency of
50 Hz or 60 Hz. Before connecting the power cord to a power source,
verify that the LINE VOLTAGE SELECTOR, located in the rear panel
power-entry module, is set so that the correct AC line voltage value
is visible.
The SR124 Analog Lock-In Amplifier will be damaged if operated
CAUTION
with the LINE VOLTAGE SELECTOR set for the wrong AC line
voltage, or if the wrong fuses are installed. Verify that the correct
line fuses are installed before connecting the line cord. Fuse size is
5MF “fast blow” (
5
×
20 mm). For 100 V
/
120 V, use 4 A fuses; for
220 V
/
240 V, use 2 A fuses.
Line cord
The SR124 Analog Lock-In Amplifier has a detachable, three-wire
power cord for connection to the power source and to a protective
ground. The chassis of the instrument is connected to the outlet
ground to protect against electrical shock. Always use an outlet
which has a properly connected protective ground.
Service
The SR124 Analog Lock-In Amplifier does not have any user service-
able parts inside. Refer service to a qualified technician.
Do not install substitute parts or perform any unauthorized modi-
fications to this instrument. Contact the factory for instructions on
how to return the instrument for authorized service and adjustment.
v