1.4
Signal and input filter
1 – 3
4. Connect a BNC cable from the front-panel Ref Out BNC (right-
hand most connector) to the A
/
I input of the Signal Input (left-
hand most connector).
5. Switch on the AC power. Allow the unit to warm up for 1 hour
for full specified performance.
1.4
Signal and input filter
1. Restore defaults:
Press [Recall],
and then turn the
REFERENCE knob (large right-hand side knob) clockwise un-
til the display shows ”deFLt”. Press [Recall] a second time to
restore factory defaults.
2. Verify the panel meter shows approximately
+
20% (positive)
deviation.
3. Turn the
SENSITIVITY knob counterclockwise 2 clicks, to se-
lect 100 mV. Verify the meter displays approximately
+
100%
(positive) deviation.
4. Change the filter to Notch by pressing the [Type] button four
times. The meter should swing to approximately 0% deviation.
5. Turn the
Q-factor knob clockwise until Q
=
100. The meter
should remain near 0% deviation.
6. Press the [Mode] button in the OUTPUT block to switch to AC
Volt mode. Turn the
SENSITIVITY knob 1 click counterclock-
wise, to select 50 mV. None of the overload indicators should
be illuminated.
7. Slowly (one click at a time) turn the large
INPUT FILTER knob
to minimize the meter display as close to 0 (no deflection) as
possible.
8. Press the
INPUT FILTER knob inward once, to illuminate the
f trim
indicator.
9. Adust the frequency fine trim by turning
INPUT FILTER to
mimimize the meter deflection (minimum is not very sensitive
to
f trim
).
10. Press the
INPUT FILTER inward once more, to illuminate the
depth
indicator.
11. Adjust the depth trim by turning
INPUT FILTER to mim-
imize the meter deflection (minimum is somewhat sensitive to
depth
).
SR124
Analog Lock-In Amplifier