7. Checking the measuring tool
7.1 Accuracy check
To prevent incorrect measurements, the accuracy
must be checked at regular intervals, e.g. before you
start work, or after hard knocks or
major changes
in
temperature.
Step 1:
Place the device with the lower measuring sole on as
horizontal a surface as possible (e.g. table) with the
display side facing the user. Determine the measurement.
Step 2:
Turn the unit by 180° in the same position.
Step 3:
The rear of the unit is now facing the user.
Compare the new measurement with the measurement
determined in step 1. In the case of deviations > 0.05°,
the unit must be calibrated again (-> Calibration).