![St. Jude Medical Tendril MRI LPA1200M Скачать руководство пользователя страница 25](http://html.mh-extra.com/html/st-jude-medical/tendril-mri-lpa1200m/tendril-mri-lpa1200m_user-manual_1356343025.webp)
23
Figure 7. Extend the helix by rotating the fixation tool clockwise
Retract the helix by holding the lead body stationary in one hand and turning only the gray portion
of the tool opposite the direction indicated by the arrow "FIXATE" (counterclockwise).
Figure 8. Retract the helix by rotating the fixation tool counterclockwise
Test the Mechanical Operation of the Helix with the Clip-On Tool
For information on using the clip-on tool, see Secure the Tip with the Clip-On Tool (page 25).
Using the Vein Lifter
A vein lifter is supplied to facilitate the introduction of the lead into a free-standing vein. Insert the
tip of the vein lifter into the vein incision and gently lift it while introducing the lead underneath,
into the vein.
Содержание Tendril MRI LPA1200M
Страница 1: ...Tendril MRI Pacing leads Model LPA1200M User s Manual ...
Страница 13: ...11 Figure 1 Lead Safety Phase The figure below depicts the MRI Phase ...
Страница 28: ...26 Figure 10 Open the clip on tool 1 Insert lead into notch ...
Страница 35: ......
Страница 36: ......