31
Technical Specifications
Table 8. Technical specifications
Length
46, 52, 58 cm
Connector Type
IS-1
bipolar
Maximum lead body size
7.9 French
Lead body tubing diameter
6.6 French
Fixation mechanism
Helix
Electrode configuration
Tip
Active helix
Ring
Cylindrical
Electrode surface area
Tip
6.0 mm
2
Rings
16.5 mm
2
Electrode length
Tip (extended)
1.8 mm
Tip to ring spacing
10 mm
Electrical resistance
Tip to connector pin
46 cm: 20
Ω
52 cm: 22
Ω
58 cm: 24
Ω
Ring to connector ring
46 cm: 46
Ω
52 cm: 52
Ω
58 cm: 59
Ω
Materials
Conductors
Multifilar MP35N
Connector
Stainless steel
Lead body
Optim
™
insulation
Tip and ring electrodes
Titanium nitride-coated platinum-iridium alloy
Soft tip
Silicone rubber
Lead tip header
PEEK, titanium
Total steroid dose -
monolithic controlled
release device (MCRD)
460 micrograms dexamethasone sodium phosphate (DSP)
MR Conditional markers
Platinum
Suture sleeve
Silicone rubber
Stylet
Stainless steel
Typical number of rotations to extend helix for initial placement
Straight stylet
6-10
J-stylet
8-11
9
St. Jude Medical IS-1 lead connectors are compatible with connector cavities that conform to the international connector
standard ISO 5841-3.
10
MP35N is a trademark of SPS Technologies.
11
Not to exceed 20 rotations.
12
Lead length, stylet configuration, and differences in anatomy may cause variations in the number of rotations required to extend
the helix.
Содержание Tendril MRI LPA1200M
Страница 1: ...Tendril MRI Pacing leads Model LPA1200M User s Manual ...
Страница 13: ...11 Figure 1 Lead Safety Phase The figure below depicts the MRI Phase ...
Страница 28: ...26 Figure 10 Open the clip on tool 1 Insert lead into notch ...
Страница 35: ......
Страница 36: ......
Страница 37: ......