Guidance and manufacturer’s declaration –
electromagnetic immunity
Immunity Test IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment
Guidance
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields
should be at
levels
characteristic of a
typical location in
a typical
commercial or
hospital
environment.
Conducted RF
IEC 61000-4-6
6 Vrms
150 kHz to
80 MHz
6 Vrms
(ISM/Radio
bands
between 150
kHz to 80
MHz)
80% AM at 1
kHz
6 Vrms
Portable and
mobile RF
communications
equipment
should be used
no closer to any
part of the
device, including
cables, than the
recommended
separation
distance
calculated from
the equation
applicable to the
frequency of the
transmitter.