Table 11.
Guidance and manufacturer’s declaration –
electromagnetic immunity
Immunity Test IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
The relative
humidity should
be at least 5%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power supply
lines
±1 kV for
input/output
lines
Not
applicable
No guidance for
battery-powered
devices.
Surge
IEC 61000-4-5
±1 kV line to
line
±2 kV line to
earth
Not
applicable
No guidance for
battery-powered
devices.
Voltage dips,
short
interruptions
IEC 61000-4-11
<5% U
T
40% U
T
70% U
T
Not
applicable
No guidance for
battery-powered
devices.