80
to mitigate this effect by reorienting or
relocating the receiving device.
CAUTION: To avoid increasing emissions
or decreasing immunity from a device or
system, use only St. Jude Medical-
approved components with this system. Do
not use St. Jude Medical components with
other non-St. Jude Medical devices or
systems.
Table 10.
Guidance and manufacturer’s declaration –
electromagnetic emissions
Emissions
Test
Compliance Electromagnetic Environment
Guidance
RF emissions
CISPR 11
Group 1
The device uses RF energy for its
internal and system interface
functions. Its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The device is suitable for use in all
establishments including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic
purposes.
Harmonic
emissions
IEC 61000-3-2
Not
applicable
Voltage
fluctuations/
flicker
emissions
IEC 61000-3-3
Not
applicable
Содержание 3875
Страница 1: ...St Jude Medical Patient Controller For Deep Brain Stimulation Systems Model 3875 User s Guide ...
Страница 36: ...32 Figure 2 Patient controller features 1 Power button 2 Patient controller Home button ...
Страница 55: ...51 Figure 10 Strength screen 1 Done button 2 Decrease button 3 Increase button ...
Страница 57: ...53 Figure 11 System screen ...
Страница 94: ......
Страница 95: ......