Single Gas Analyzer (SGA)
Operations Manual
Super Systems Inc.
Page 33 of 44
Page 33 of 51
Modbus Registers
The following table shows the Modbus registers for the sensor module. The name of the
register, address location, and description are provided.
Register Name
Register Location
Description
VERSION_NUMBER
0
current version number of the
firmware
UART_1_MODE
1
0 = slave, 1 = Sensor Driver
UART_1_BAUD_RATE
2
Baud Rate:
0=1200,...,5=19200,...10=115200.
UART_2_MODE
3
0 = slave, 1 = Sensor Driver
UART_2_BAUD_RATE
4
Baud Rate:
0=1200,...,5=19200,...10=115200.
UART_3_MODE
5
0 = slave, 1 = Sensor Driver
UART_3_BAUD_RATE
6
Baud Rate:
0=1200,...,5=19200,...10=115200.
UART_4_MODE
7
0 = slave, 1 = Sensor Driver
UART_4_BAUD_RATE
8
Baud Rate:
0=1200,...,5=19200,...10=115200.
UART_5_MODE
9
0 = slave, 1 = Sensor Driver
UART_5_BAUD_RATE
10
Baud Rate:
0=1200,...,5=19200,...10=115200.
BOARD_ADDR
14
Board modbus address
(important for slave only)
MODEL_NUM
15
MODEL number Map as reg 900
SET_FACT_DEF
16
23205 = Full Defaults, 23206 =
H2 Defaults, 23207 = Loop 1
Defaults, 23208 = Loop 2
Defaults
DEGREE_REG
17
0 = °F, 1 = °C, 2 = °R, 3 = K
CUR_LOOP_CAL_REG
18
Calibration state. 0 = normal, 1
= prep zero, 2 = store zero, 3 =
prep span, 4 = store span
CUR_LOOP_CAL_CHN
19
Calibration channel
CUR_LOOP_CAL_VAL
20
Cal value. 20.12 mA would be
20120