background image

MODEL SR465 

FLOOR SCALE SYSTEM OPERATION

INITIALIZING THE SYSTEM ZERO:

Before the patient is positioned on to the scale system you must first set the system zero. This

is done as follows;

1. Press the 

"O"

 button. The displayed message will indicate 

" HANDS OFF "

. Make

sure that nothing is in contact with the scale system while zeroing. In a few seconds

the display will indicate 

"WEIGHT =  0.0 LB "

.

Note: When the unit is first turned on the mode will default to the last mode used.

The patient may now be positioned on to the weighing surface to view the weight data.

BUTTON FUNCTIONS :

The 

"0"

 button is used to initialize the system zero before placing a patient on to the

scale system. When the 

"0"

 button is pressed, the display message will indicate

"HANDS OFF".

 At this point you should make sure that nothing is in contact with the

weighing surface so that the system zero will be accurate. In a few moments the

display will read 

- "WEIGHT =  0.0 Lb"

.

The display mode will automatically default to the last mode used

If after positioning the patient on to the scale system, the display shuts off before

you are able to view the weight data, simply press the 

"WEIGH"

 button to WAKE the

display and view the patient's weight.

NOTE: It is recommended that you ZERO the system before each weighing to
assure accuracy.

The 

"HOLD"

 button will freeze the displayed patient weight and store it away in

memory. To recall the LAST WEIGHT after the system has shut off, press the

"HOLD"

 button BEFORE pressing any other. The system only recalls the LAST

WEIGHT that was stored by HOLD.

Weight data may be viewed in either Pounds or Kilograms. Pressing the 

"MODE"

button allows you to toggle the POUNDS and KILOGRAMS display modes. Both
modes will display a resolution of 0.1 pounds or kilograms.

page 2

Part No. MAN465

                                                                                                                                                           rev: 000706

ZERO

0

WEIGH

MODE

LB

KG

HOLD

ZERO

0

Содержание SRScales SR465

Страница 1: ...S S by SR Instruments Inc Model SR465 In Floor Platform Scale System OWNERS MANUAL SR Instruments Inc 600 Young Street Tonawanda NY 14150 Tel 888 WEIGH ME Fax 716 693 5854 Part No MAN465 rev 000706 ...

Страница 2: ...ions Page 1 System Operation and button Functions Page 2 SECTION II TECHNICAL SERVICE INFORMATION Calibration and Maintenance Page 4 Care and Cleaning Page 5 Schematic Diagram Page 6 Part List and Component Layout Page 7 Platform Installation Instructions Page 8 Part No MAN465 rev 000706 ...

Страница 3: ...from six 6 common D cell alkaline batteries which will provide over 20 000 weight reading before needing replacement This eliminates the need for an external battery charger or an A C outlet The Patient s weight is displayed on a user friendly 16 character dot matrix liquid crystal display This display uses bright green L E D back lighting for superior visibility in all lighting conditions With th...

Страница 4: ...with the weighing surface so that the system zero will be accurate In a few moments the display will read WEIGHT 0 0 Lb The display mode will automatically default to the last mode used If after positioning the patient on to the scale system the display shuts off before you are able to view the weight data simply press the WEIGH button to WAKE the display and view the patient s weight NOTE It is r...

Страница 5: ...SECTION II TECHNICAL SERVICE INFORMATION For Sales Product Information and Technical Assistance Please Call 1 800 654 6360 Part No MAN465 rev 000706 ...

Страница 6: ... in the CAL MODE press the ZERO button to zero the display 4 Place a known calibrated weight on to the weighing surface and compare it to the displayed reading 5 Use the LEFT CAL button to make corrections and the RIGHT button to make corrections to the displayed weight The displayed value should be within 0 1 of the calibrated weight plus or minus 1 digit of reading When adjustments are completed...

Страница 7: ...mend mild soap and water for general cleaning The floor transducer is covered by a Stainless Steel plate and may be vacuumed or swept BATTERY REPLACEMENT model SR465 only When battery replacement is necessary the display will indicate REPLACE BATTERY To replace the batteries 1 Remove the instruction cover plate to expose the battery compartment 2 Replace ALL SIX 6 Batteries as shown 3 Replace the ...

Страница 8: ...SCHEMATIC and COMPONENT LAYOUT Part No MAN465 rev 000706 ...

Страница 9: ...SISTOR VOLTAGE REF ZVNL110 1 Q1 SE1857D255 ET2N3906 TRANSISTOR 2N3906 MATOROLA 1 Q2 SE1857D255 ER17505W3 75 OHM 3 WATT 5 RESISTOR 1 R2 SE1857D255 EV5KKVSF82 5 K POTENTIOMETER 1 R1 SE1857D255 ER41005QF 10K 1 4 WATT 5 RESISTOR 2 R3 R4 SE1857D255 ER62001EF 2MEG 1 RES 1 R5 SE1857D255 ER57503EJ 750K 1 8 WATT 1 J RESISTOR 1 R12 SE1857D255 ER44531EJ 45 3K 1 8 WATT 1 J RESISTOR 2 R6 R8 SE1857D255 ER43241E...

Страница 10: ...er bolts A so that the platform base is one inch 1 below the surface of the floor on all sides This will ensure that the deck is even with the floor surface when installed NOTE If you are planning on tiling or carpeting the deck be sure to compensate for the thickness of material You may place the deck on the platform to check the level as you make your adjustments 9 When the 4 corners are done ad...

Страница 11: ...Part No MAN465 rev 000706 ...

Страница 12: ...Part No MAN465 rev 000706 ...

Отзывы: