User Manual, Version 01/2004
11
2.4 Delivery Inspections
On delivery, inspect the unit to make sure that it has not been damaged during transport and that all
accessories are intact and complete (see Section 2.1). Inform the supplier immediately in the event of any
damage or missing items. Do not use the unit if it has been damaged.
2.5 Mains Voltage
Your unit is suitable for use either with a mains voltage of 230 V, 50-60 Hz
or
115 V, 50-60 Hz. Check the
label and the back of the control panel to make sure the rated voltage of the unit matches that of your
region (connecting the unit to a higher voltage can cause irreparable damage to the unit).
2.6 Attaching the Support to the Vibration Plate (Diagram 1)
The handrail must be fitted to the vibration plate before using the "fitvibe medical Pro" unit.
We recommend that two people carry out installation of the “fitvibe medical Pro”. The SW 17 socket
spanner (30) as well as the enclosed hexagonal nuts (31) and spring washers (32) should be used during
installation.
Before attaching the handrail, position the vibration plate in the location where it will be used. The unit
must be placed on a solid concrete floor.
Now attach the handrail to the vibration plate as follows:
Tilt the plate forwards so that it rests securely on its long narrow front edge. Place a chair approximately 1
meter away from the plate. Attach the handrail support to the four bolts on the rear of the vibration plate.
Rest the upper part of the handrail on the chair to hold it in place (see Diagram 1).
Attach a spring washer and hexagonal nut to each of the bolts in turn. Simply slide a washer over each of
the bolts. When fitting the hexagonal bolts, it is helpful to place a nut in the socket spanner and then
place it on the bolt. Turn the spanner clockwise to tighten the nut. Repeat this process for each bolt.
Tighten the hexagonal nuts sufficiently to ensure the handrail is securely attached to the vibration plate.
Do not tighten them fully. (Note: The hexagonal nuts are self-locking to prevent them from being loosened
by vibration. Some light resistance can be felt when tightening them.)
Place the vibration plate upright and adjust the handrail so that it is perpendicular to the plate. Now fully
tighten the hexagonal nuts.
2.7 Connecting and Switching On the "fitvibe medical Pro" Vibration Unit (Diagram 2)
Use the plug (22) to connect the cable (23) leading from the vibration plate (1) to the control panel (6).
Make sure that the small ridge on the plug is facing downwards before attempting to connect it to the
control panel. Tighten the plug fully. Now connect the hand-held switch (16) to the control panel.
The power input (20) is located on the bottom of the control panel towards the back. Use the main switch
to switch on the “fitvibe medical Pro” vibration unit.
A welcome message appears on the display.
In the lower right corner of the screen can be seen the software version number .
Simply press any key to go to the main menu.
2.8 System Settings
The "fitvibe medical Pro" unit features a multiplicity of system settings which allow you to finely tune the
unit to your own specific requirements. However, when you start the unit for the first time you should take
some time to study the pre-programmed settings before making any necessary changes.
To access the system settings menu from the main menu,
press the system settings button:
Using the combined rotary and push button (11), you can select the various menu options.