3. INSULATION RESISTANCE MEASUREMENTS 

a)

Turn the function switch from the 

OFF

 position to the left

4000 

/1000V---4000/ 

500V---4000 

/250V---1000 

/125V

,and chose one of the voltage-block( there are 4 ranges 

namely, 4 

40 

400 

4000 

,  can be switched 

automatically for every voltage-block.) 

b)

Connect two testing lines to the tested

 

c)

Push down and hold the 

TEST

  button /or press the 

  LOCK

” 

keystoke  first and then the 

TEST

  button, if the tested is 

electriferous  and its voltage ( AC/DC) is over 30V, it will refuse 
work and no high-voltage testing  occurs,  simultaneity, it shows 

>30V

  on the LCD, the symbol 

      

  flashes, and the buzzer 

warns frequently. if the tested is diselectriferous or its voltage is 
lower than 30V, it will enter into the formal testing process and 
brings  the high-voltage.  on  the  primary  display,  the insulation 
resistance in M

Ω

  is indicated  in-phase  with analog bar;  on  the 

secondary  display,  the tested insulation voltage in V (DC)  is 
indicated, the symbol 

      

  flashes and the buzzer  warns 

frequently 

d)

Being free from the 

TEST

  button  or pushing  down the 

TEST

 

button in the 

LOCK 

 status can exit from the 

LOCK

 status and 

shutoff the high-voltage

synchronously, the resistance values is 

indicated in the primary  display will be held, and the secondary 
display still be in the status of monitoring the insulation voltage for 
the tested. 

e)

Subsequently, discharge the balance insulation voltage of the 

tested through the inner switch of the meter. 

  Turning the function switch can exit automatically from testing 

status during the process

 

Содержание 420001

Страница 1: ...Electrical Insulation Tester 420001...

Страница 2: ...working around bare conductors or bus bars Accidental contact with the conductor could result in electric shock Use the meter only as specified in this manual otherwise the protection provided by the...

Страница 3: ...Environment conditions Installation Categories Pollution Degree 2 Altitude up to 2000 meters Indoor use only Relatively humidity 80 max Operation Ambient 0 40 C Maintenance Clearing Repairs or servic...

Страница 4: ...14 F to 140 F and Humidity below 70 RH Power source DC9V 6x1 5V Size AA battery or Equivalent Dimensions 200 L x 92 W x 50 H mm Weight Approx 700g include battery Accessories Test leads 6pcs battery...

Страница 5: ...1 2 10 10M 750Vrms Meg OHMS Terminal Voltage Range Resoluti on Accuracy Test Current Short circuit current 125V 0 10 0 125 4 000 M 0 001M 2 10 1mA load 125k 1mA 4 001 40 00 M 0 01M 2 10 40 01 400 0 M...

Страница 6: ...test leads On M Range and 400 BZ ACV DCV Connect the red test lead into the V terminal and the black lead into the COM terminal 2 2 Battery Check UP Replacement a As battery power is not sufficient LC...

Страница 7: ...o the MIN if instant pressing once more it will recycle but exit if pressing 2 seconds again LOCK In the insulation resistance testing function press the LOCK button and then push down the TEST key it...

Страница 8: ...nous with the primary display While testing the insulation resistance the symbol flashes frequently if the voltage is over 30V While testing the insulation resistance the symbol flashes frequently and...

Страница 9: ...erous or its voltage is lower than 30V it will enter into the formal testing process and brings the high voltage on the primary display the insulation resistance in M is indicated in phase with analog...

Страница 10: ...us with the analog bar d When the impedance on circuit is below approximately 35 It will indicate by a continuous beeper e The current is from 200 to 220mA while the tested resistance is 0 f The high...

Страница 11: ...terminals or by opening the main switch If the main switch is used and the motor also has a starter then the starter must be held by some means in the ON position In the latter case the measured resis...

Страница 12: ...armature field coils and brush rigging by connecting one megohmmeter lead to each of them individually leaving the other connected to the grounded motor housing The above also applies to DC Gemerator...

Страница 13: ...from other equipment Check each conductor to ground and or lead sheath by connecting one megohmmeter lead to a ground and or lead sheather and the other megohmmeter lead to each of the conductors in t...

Страница 14: ...ranty does not cover probes batteries battery leakage or damage resulting from accident tampering misuse or abuse of the product Opening the meter to expose its electronics will break the waterproof s...

Отзывы: