PAGE 3 OF 4
S P E L L M A N H I G H V O LTA G E E L E C T R O N I C S C O R P O R A T I O N
XRV
1.8-6.0KW X-RAY
GENERATORS
USA
+1-631-630-3000
FAX: +1-631-435-1620
UK
+44 (0)1798 877000
FAX: +44 (0)1798 872479
JAPAN
+81 (0)48-447-6500
FAX: +81 (0)48-447-6501
CHINA
+86 (0)512-67630010
FAX: +86 (0)512-67630030
e-mail: [email protected]
www.spellmanhv.com
128060-001
REV.F
Spellman High Voltage is an ISO 9001:2000 and ISO 14001:2004 registered company
RS-232 DIGITAL INTERFACE—
J3 9 PIN FEMALE D CONNECTOR
PIN
SIGNAL
PARAMETERS
1
NC
No Connection
2
TX out
Receive Data
3
RX in
Transmit Data
4
NC
No Connection
5
SGND
Ground
6
NC
No Connection
7
NC
No Connection
8
NC
No Connection
9
NC
No Connection
J1 HV CONNECTOR—R24/R28
PIN
SIGNAL
PARAMETERS
C
HV Output
XRV160 and XRV320—R24 Connector
XRV225 and XRV450—R28 Connector
S
Small Filament Output
0 to 6 amps @ 10Vdc
L
Large Filament Output
0 to 6 amps @ 10Vdc
J2 ANALOG INTERFACE—25 PIN D CONNECTOR
PIN
SIGNAL
PARAMETERS
1
Power Supply Fault
Low, sum of faults, HVPS detected a fault,
open collector, 50V @ 10mA max
2
mA Program
0 to 10V FS Z in = 10M ohms
3
kV Program
0 to 10V FS Z in = 10M ohms
4
Filament Limit L/S Ref.*
0 to 10V FS Z in = 10M ohms
5
Filament Preheat L/S Ref.* 0 to 10V FS Z in = 10M ohms
6
kV Monitor
0 to 10V FS Z out = 4.99k ohms
7
mA Monitor
0 to 10V FS Z out = 4.99k ohms
8
Filament Current Monitor*
0 to 10V FS Z out = 4.99k ohms
9
Signal Ground
Ground
10
X-Ray Enable
+24Vdc = X-Ray ON, connect to pin 14 with
dry contact relay
11
Filament ON*
Filament ON status, low, filament is ON
open collector 50V, @ 10mA max
12
Interlock 1
Active low, interlock is closed, safe to enable HV
13
Interlock 2
Active low, interlock is closed, safe to enable HV
14
+24Vdc
+24Vdc @ 100mA, maximum
15
Filament Enable*
Active low, turn filament ON
16
Filament Control*
Active low, filament is regulated by ECR (HV must
be ON). Not active, the filament is regulated by the
preheat reference
17
Filament L/S Select
Filament selection large or small,
low = small spot is selected
18
Filament L/S Confirm
Open collector, 50V @ 10mA max
Filament selection confirm, low = small spot is selected
19
HVPS RDY
Low = HVPS ready, open collector, 50V @ 10mA max
20
X-Ray ON
X-Ray ON status, low = X-Rays are ON
open collector, 50V @ 10mA max
21
Interlock Status
Low, interlocks are closed, can enable HV
open collector, 50V @ 10mA max
22
GND
Digital ground
23
X-Ray ON Pre-Warn
Pre-warning, low, before X-Ray ON
open collector, 50V @ 10mA max
24
Reset
Active low, minimum 10mS transition
25
Arc fault
Low, arc fault, the HVPS has detected an arc
open collector, 50V @ 10mA max
ETHERNET DIGITAL INTERFACE—
J4 8 PIN RJ45 CONNECTOR
PIN
SIGNAL
PARAMETERS
1
TX+
Transmit Data +
2
TX-
Transmit Data -
3
RX+
Receive Data +
4
NC
No Connection
5
NC
No Connection
6
RX-
Receive Data -
7
NC
No Connection
8
NC
No Connection
USB DIGITAL INTERFACE—
J5 4 PIN USB “B” CONNECTOR
PIN
SIGNAL
PARAMETERS
1
VBUS
+5 Vdc
2
D-
Data -
3
D+
Data +
4
GND
Ground
JB1 MAIN AND AUXILIARY INPUT POWER—
TYPE 97-3102A-24-11P (Single Phase Units)
PIN
SIGNAL
PARAMETERS
A
Auxiliary AC Line Power
180-264Vac
B
Auxiliary Ground
Ground
C
Auxiliary AC Neutral
Neutral
D
Main AC Line Power
180-264Vac
E
Main Ground
Ground
F
Main AC Neutral
Neutral
*Not active on positive models
MODEL SELECTION TABLE
MODEL
VOLTAGE
POWER
POLARITY
XRV160*1800
160kV
1.8kW
Pos or Neg
XRV160*3000
160kV
3.0kW
Pos or Neg
XRV160*4000
160kV
4.0kW
Pos or Neg
XRV160*6000
160kV
6.0kW
Pos or Neg
XRV225*1800
225kV
1.8kW
Pos or Neg
XRV225*3000
225kV
3.0kW
Pos or Neg
XRV225*4000
225kV
4.0kW
Pos or Neg
XRV320P&N1800
±160kV
1.8kW
Bipolar
XRV320P&N4500
±160kV
4.5kW
Bipolar
XRV450P&N1800
±225kV
1.8kW
Bipolar
XRV450P&N4500
±225kV
4.5kW
Bipolar
*Specify P for positive polarity and N for negative polarity. Positive polarity models do not have
integrated filament power supplies. Contact Spellman for custom output voltage/power models.
JB1 MAIN AC INPUT POWER—
TYPE 97-3102A-24-22P (Three Phase Units)
PIN
SIGNAL
PARAMETERS
A
Line 1
208Vac, ±10%, 50/60Hz @ 25 amps
B
Line 2
208Vac, ±10%, 50/60Hz @ 25 amps
C
Line 3
208Vac, ±10%, 50/60Hz @ 25 amps
D
GND
Ground
JB2 AUXILIARY AC INPUT POWER—
TYPE 97-3102A-20-3P (Three Phase Units)
PIN
SIGNAL
PARAMETERS
A
Line 1
208Vac, ±10%, 50/60Hz (source 3 phase L1, L2)
B
Line 2
208Vac, ±10%, 50/60Hz (source 3 phase L1, L2)
C
GND
Ground
Note: Use 4 conductor cable or single isolated wires rated no
less that 600Vac, 30 amps (10AWG, minimum)
System Ground: System ground wires (10AWG minimum) to
the ground terminal E1 GND to the power supply using ground