X2364 MANUAL
23
118044-001 REV E
The fault signals that are checked are:
Cooling
Fault
Overcurrent
Fault
Overpower
Fault
Overvoltage
Fault
Arc
Fault
Regulation Error Fault
Interlock Status
(open is interpreted as a fault)
8.5 Command Summary
Commands sent from the computer to the power supply
start with the “SOH” character followed by a command
identifier letter. The defined commands from the
computer to the power supply are:
S - Set Power Supply
(contains desired power supply
settings) – 18 bytes.
Q - Query Power Supply
(requests that the power
supply returns to the computer a Response Packet) – 5
bytes.
V - Request Power Supply Software Version Level
–
5 bytes.
Responses from the power supply back to the computer
start with an A, R, E or B character:
A – Acknowledge Response Packet – 2 bytes
R – Data Response Packet – 16 bytes
E – Error Response Packet – 5 bytes
B – Software Version Level Response Packet – 6
bytes
All command and response packets end with a carriage
return character <CR>.