X2364 MANUAL
8
118044-001 REV E
1
TB1
Install jumper
2
3
4
if COOLING FAULT
INTERLOCK 2
6
7
5
9
10
8
Cooling FAULT
Isolated contacts;
must be closed to operate
"HV ON" Fail-safe Lamp
12V, 120mA. Do NOT Short!
Isolated contacts;
must be closed to operate
INTERLOCK 1
Isolated contacts;
must be closed to operate
+15Vdc Return
+15Vdc, 0 to 50mA max.
switch is not used
Do Not Short!
Install proper resistor if
FAIL SAFE lamp is not used
(100 ohms, 2W or higher)
Install jumper
if INTERLOCK contacts
are not used
Install jumper
if INTERLOCK contacts
are not used
Signals are ~ 15Vdc, 25mA max and are only to be used for contact closure.
They should be located as close as possible to the power supply
It is recommended to use isolated contacts / switches or relays for both interlocks.
Figure 3.2 Rear Panel I/O Interface