600 SERIES MANUAL
2
118118-001 Rev A
SECTION I –
INTRODUCTION/SPECIFICATIONS
1.0 SCOPE
OF
MANUAL
This manual is provided to assist the user in
the installation and operation of the Bertan
Series 602C, 603C, 605C, 606C and 612C
modular high voltage power supplies.
Statements will apply to models in all of the
Series unless reference is made to specific
models. For the protection of personnel and
equipment, it is essential that this manual
be thoroughly read prior to the installation
and application of power.
1.1 PURPOSE
OF
EQUIPMENT
The Series 602C, 603C, 605C, 606C and
612C are families of regulated fixed output
polarity modular high voltage power
supplies. They provide exceptional
performance in applications such as CRT’s
spectrometers, detectors, medical imaging,
image intensifiers, E-Beam, I-Beam,
medical imaging and capillary
electrophoresis (HPCE) systems.
1.2 DESCRIPTION
The units are fully enclosed and designed to
easily satisfy system or bench top
operation. A wide range of stable output
voltages, up to 30kV are available. The
output voltage is controlled locally by a
minimum 15-turn potentiometer. Remote
analog voltage of resistance programming is
also available to the user as a standard
feature. All units offer a 0 to +5Vdc analog
monitor output proportional to the output
current.
The Series 602C, 603C and 612C require
an input of 115/230Vac (switch selectable)
+10%, 50-60Hz. The Series 605C and 606C
require a +28Vdc +10% input (or +24Vdc
±10% for units equipped with +24V Option).
Each unit in the Series converts the applied
power to a high voltage DC output. This
output voltage is highly regulated and
filtered. The high voltage assembly is fully
encapsulated in silicone rubber for reliable,
arc-free, stable operation.
HV Connector:
SERIES 602C, 603C, 605C
606C
MODEL
HV OUTPUT
CONNECTOR
MATING HV
CONNECTOR
602C
-15 P, N thru -50 P, N
UG-931/U (MHV)
UG-932/U (Bertan PDB)
603C & 606C
-15 P, N thru -50 P, N
KINGS 1707-1 (SHV)
1705-1
Bertan PAE)
All 605C and all
-100 P, N and -150 P, N
8101FP
8101M
(Bertan PGC-008L151-000)
All –200P,N
8101FP
8101M
(Bertan PGC-008L1 51-000)
All –300P,N
8111SFP
8111M
(Bertan PGQ-008L201 -000)
Only the mating connectors for the 8101 FP
and 8111 SFP (assembled to an unshielded
high voltage cable) are provided. For other HV
mating connectors order separately.