Trigger modes and appendant registers
Main analog external trigger (Ext0)
(c) Spectrum Instrumentation GmbH
105
Trigger on positive and negative edge
The trigger input is continuously sampled with the selected
sample rate. If the programmed trigger level is crossed by
the trigger signal (either rising or falling edge) the trigger
event will be detected.
This edge triggered external trigger mode correspond to
the trigger possibilities of usual oscilloscopes.
High level trigger
This trigger mode will generate an internal gate signal that
can be useful in conjunction with a second trigger mode to
gate that second trigger. If using this mode as a single trigger
source the card will detect a trigger event at the time when
entering the high level (acting like positive edge trigger) or if
the trigger signal is already above the programmed level at
the start it will immediately detect a trigger event.
The trigger input is continuously sampled with the selected
sample rate. The trigger event will be detected if the trigger
input is above the programmed trigger level.
Low level trigger
This trigger mode will generate an internal gate signal that
can be useful in conjunction with a second trigger mode to
gate that second trigger. If using this mode as a single trigger
source the card will detect a trigger event at the time when
entering the low level (acting like negative edge trigger) or if
the trigger signal is already above the programmed level at
the start it will immediately detect a trigger event.
The trigger input is continuously sampled with the selected
sample rate. The trigger event will be detected if the trigger
input is below the programmed trigger level.
Register
Value
Direction
set to
Value
SPC_TRIG_EXT0_MODE
40510
read/write
SPC_TM_BOTH
4h
SPC_TRIG_EXT0_LEVEL0
42320
read/write
Set it to the desired trigger level in mV
mV
Register
Value
Direction
set to
Value
SPC_TRIG_EXT0_MODE
40510
read/write
SPC_TM_HIGH
00000008h
SPC_TRIG_EXT0_LEVEL0
42320
read/write
Set it to the upper trigger level in mV
mV
Register
Value
Direction
set to
Value
SPC_TRIG_EXT0_MODE
40510
read/write
SPC_TM_LOW
00000010h
SPC_TRIG_EXT0_LEVEL0
42320
read/write
Set it to the upper trigger level in mV