SERVOPRO PureGas Gas Purifier
02005001A / Revision 0
12
4.2
Mechanical Installation and purging
The purifier comes sealed in an argon atmosphere and must be purged before
purification can begin.
Purge the lines, with an inert gas, upstream of the purifier thoroughly.
While under purge, using standard UHP procedures, remove the inlet cap and
connect the inlet fitting to the line as quickly as possible, no longer than 30 seconds.
Remove the outlet fitting cap and, using standard UHP procedures, connect the
downstream gas line as quickly as possible, no longer than 30 seconds.
The purifier should be purged with the application gas for at least ten minutes prior to
the start of heating.
The purifier can now be activated.
WARNING
Do not remove any fittings or open any valves until the inlet and outlet
lines are connected and purged.
WARNING
The purging process must be carried out at room temperature. Do not
switch the purifier on at this stage.
WARNING
Ensure that the pipes to be connected to the purifier are routed so that
they do not present a trip hazard to people.
WARNING
Purge and sample gases are asphyxiant in high concentrations. Ensure
that the purge outlet pipe is vented to an area where the gases will not be
a suffocation hazard to people.
WARNING
Purge and sample gases are asphyxiant in high concentrations. To
prevent the build-up of such gases, ensure that the purifier is used in a
sufficiently well-ventilated environment.
WARNING
Strict adherence to UHP components handling procedures is required to
maintain the factory specifications. Do not allow the purifier column and
the internal piping surface to be exposed to atmosphere. The purifier
should be purged with the application gas only.
Содержание Servomex Servopro PureGas
Страница 1: ...Part Number 02005001A Revision 0 Language UK English SERVOPRO PureGas OPERATOR MANUAL OPERATOR MANUAL ...
Страница 2: ...This page is intentionally blank ...
Страница 11: ...SERVOPRO PureGas Gas Purifier 02005001A Revision 0 10 4 1 1 Figure 1 Overall and fixing dimensions Standard ...
Страница 12: ...SERVOPRO PureGas Gas Purifier 02005001A Revision 0 11 4 1 2 Figure 2 Overall and fixing dimensions VCR ...