6 Disconnecting the Purifier
If it is necessary to disconnect the purifier to carry out plant/factory maintenance, for example,
and the purifier will not be used for several days, the following procedure must be followed:
Prior to removal, the power must be shut off and the purifier allowed to cool to room
temperature (this takes approximately 3 hours).
While cooling, purge the purifier with inert gas.
While under purge, quickly disconnect the outlet line and replace the outlet cap.
Next, quickly disconnect the inlet line and replace the inlet cap.
Remove the purifier from the mounting bracket.
Upon replacement, follow the mounting, purging and activating instructions.
WARNING
Do not attempt to disconnect any of the components while the purifier is
in a heated (greater than 50
O
C) condition. Exposure of the getter alloy to
atmosphere under these circumstances could be hazardous.