RO Single-Membrane Replacement

Maintenance Regime:

  

As needed. Refer to the procedure for testing mem-

brane quality on page 19.

Materials Required:

  

Replacement Encapsulated RO Membrane

Procedure:

 

1.  Turn off water supply to the system.

 

2.  Mark, or otherwise note, the locations of the yellow, blue, and

 

 

black tubing connected to the encapsulated RO membrane

  (see 

Fig. 

C).

 

3.  Depress the collar on the membrane feed push-fitting with your

 

 

thumbnail and pull the black feed tubing from the fitting. This is the 

 

 

line that runs between the carbon filter and the encapsulated RO

  membrane. 
 

4.  Lift the encapsulated RO membrane from the holding clips.

 

5.  Remove the yellow and blue tubing from the encapsulated RO 

 

 

membrane by depressing their push-fittings with your thumbnail

 

 

and pulling the tubing from the fittings.

 

6.  Remove the flow restrictor assembly from the end of the yellow 

 

 

tubing as described in the Removal, Adjustment, and Replace

 

 

ment section on page 7.

 

7.  Place the flow restrictor in a safe location where it will not be 

 

 

accidentally crushed or damaged.

 

8.  Re-insert the blue and yellow tubing into their respective push-

 

 

fittings on your replacement encapsulated membrane, as described 

 

 

in the Removal, Adjustment, and Replacement section on page 

 

 

7. Snap the membrane assembly back into its holding clips.

System Maintenance

RO Membrane Replacement

If you have a Single-Membrane System (Fig. A), follow the direc-
tions on this page. If you have a Dual-Membrane System (Fig. B), 
follow the directions on page 16.

SpectraPure

®

SpectraPure®Inc

 Fax  480.894.6109 Fax us toll-free 1.877.527.7873

E-mail: [email protected] Visit us on the web www.spectrapure.com

11

Notes

Sediment Pre-Filter Replacement

Maintenance Regime:

 

 For maximum contaminant removal and long mem-

brane life, the sediment and carbon pre-filters must be changed at 6-month 
intervals. If your water contains a great deal of sediment or chlorine, the pre-filters 
may require more frequent changes to maintain adequate production rate and 
extended membrane life. 

Materials Required: 

 

1 micron MicroTec™  Sediment Filter (SF-MT-1-10)

Procedure:

 

1.  Turn off water supply to the system.

 

2.  Refer to photo on page 4. Using the provided filter wrench,

 

 

remove the left clear sediment pre-filter housing. Unscrew it 

 

 

counterclockwise as viewed from the bottom

 

3.  Remove the old filter from the housing and discard.

 

4.  Thoroughly wash the housing with a mixture of hot soapy water

 

 

and a few teaspoons of household bleach. Rinse well with clean 

  hot 

water.

 

5.  Install the new pre-filter into the housing, screw the housing

 

 

back onto the assembly, and tighten securely using the supplied

  filter 

wrench.

 

6.  Proceed with carbon block filter replacement.

Содержание The Eliminator DI

Страница 1: ...f membrane elements for one year from date of receipt by the buyer providing that the loss of performance was not caused by fouling neglect or water conditions exceeding the feed water parameters list...

Страница 2: ...ane 19 20 Membrane Testing System Troubleshooting Guide 21 Membrane Troubleshooting Guide 21 System Care Tips for Long Membrane Life 22 Storage 22 Choosing a Mounting Location 22 System Components Rep...

Страница 3: ...Temperature Extremes 1 The unit must be kept out of areas that are subject to freezing temperatures 2 High temperatures greater than 100 F 38 C must be avoided If the unit is used outside avoid puttin...

Страница 4: ...Low water temperature Heat feed water or use higher GPD membrane Low feed pressure Use booster pump or use higher GPD membrane Fouled membrane Replace membrane Membrane Troubleshooting Guide The follo...

Страница 5: ...3 X Y 143 ppm 4 X Y X 143 150 0 953 5 Rejection X Y X 100 0 953 100 95 3 Membrane Hardness Rejection 95 3 Membrane OK Production Rate SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 87...

Страница 6: ...illiliters per Minute 5 If the actual concentrate flow rate is less than the optimum concentrate flow rate you must reduce the length of the flow restrictor capillary tubing as described on the next p...

Страница 7: ...ncentrate tubing Fig C for Single Membrane System or Fig D for Dual Membrane System Remove the tubing from its push fitting at the membrane as follows a Firmly depress and hold the push fitting collar...

Страница 8: ...ail and remove the blue product water inter connect tubing 9 Lift the dual RO membrane assembly from the holding clips 10 Separate the two membranes by depressing the collar on the push fitting of the...

Страница 9: ...mbrane 1 Membrane 2 feed push fitting black inter connect feed tubing blue product water interconnect tubing Membrane 1 concentrate push fitting Membrane 1 product water push fitting SpectraPure Spect...

Страница 10: ...raPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 15 Note The flow restrictor assembly should remain uninst...

Страница 11: ...as described in the Removal Adjustment and Replacement section on page 7 Snap the membrane assembly back into its holding clips System Maintenance RO Membrane Replacement If you have a Single Membran...

Страница 12: ...and check for leaks System Maintenance SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 13 System Mai...

Отзывы: