background image

SpectraPure

SpectraPure®Inc.

  

480.894.5437   Call us toll-free  1.800.685.2783    

215 South Industrial Drive,   Suite 2A   Tempe, Arizona   85281-2941

®

12

Carbon Block Filter Replacement

Maintenance Regime:

  

 At least every 6 months or when chlorine break-

through greater than 0.1 ppm occurs in the concentrate from yellow line. To 
test for chlorine breakthrough , collect a 10 ml sample of the concentrate from 
the yellow tubing and test the chlorine concentration using test kit TK-CL-25. If 
the chlorine concentration is above 0.1ppm, replace the carbon pre-filter.

Materials Required:  

1 micron Carbon Block Filter (CF-1-10)

Procedure:

 

1.  Turn off water supply to the system.

 

2.  Refer to photo on page 4. Using the provided filter

 

 

wrench, remove the right clear carbon block filter 

 

 

housing. Unscrew it counterclockwise as viewed from 

  the 

bottom.

 

3.  Remove the old filter from the housing and discard.

 

4.  Thoroughly wash the housing with a mixture of hot

 

 

soapy water and a few teaspoons of household bleach. 

 

 

Rinse well with clean hot water.

 

5.  Install the new carbon block filter, making sure that 

 

 

the black gaskets on both ends of the filter are firmly 

 

 

seated in the gasket recesses.

 

6.  Screw the housing back onto the assembly, and tighten 

 

 

securely using the supplied filter wrench.

 

7.  Turn on system water supply and check for leaks.

System Maintenance

SpectraPure

®

SpectraPure®Inc

 Fax  480.894.6109 Fax us toll-free 1.877.527.7873

E-mail: [email protected] Visit us on the web www.spectrapure.com

13

System Maintenance

Deionization Cartridge Diagnostic

The DI stage is an anion cartridge #DI-SB-CI-10. In order to determine 
the condition of the cartridge, watch for the color to change from 
the bottom to the top. The color will change from a very dark blue to 
a lighter blue or tan color. Once the cartridge has changed color by 
70 to 80%  it 

must

 be replaced. 

Failure to replace the DI cartridge 

upon exhaustion will contaminate the product water.

Note: If the pH of your tap water is too high, the resin may not 
change color. 

Deionization Cartridge Replacement 

Materials Needed:

 One DI-SB-CI-10 Deionizaton cartridge, 

     filter 

wrench.

Procedure:

1.  Remove the filter housing from its cap by unscrewing it counter
 

clockwise as viewed from the bottom.

2.  Remove and discard the old cartridge from the housing.
3.  Thoroughly wash out the housings with hot soapy water to 
 

which a few teaspoons of household bleach have been added. 

 

Rinse well with clean hot water.

4.  Install the new deionization cartridge. Make sure the cartridge is 
 

installed in the correct direction as marked on the filter housing 

 

and that the top seal is securely attached to the top of the 

 cartridge
5.  Re-install the bottom housing onto the cap by rotating it clock-
 

wise and tighten with the wrench.

6.  Turn on system and check for leaks.

Содержание The Eliminator DI

Страница 1: ...f membrane elements for one year from date of receipt by the buyer providing that the loss of performance was not caused by fouling neglect or water conditions exceeding the feed water parameters list...

Страница 2: ...ane 19 20 Membrane Testing System Troubleshooting Guide 21 Membrane Troubleshooting Guide 21 System Care Tips for Long Membrane Life 22 Storage 22 Choosing a Mounting Location 22 System Components Rep...

Страница 3: ...Temperature Extremes 1 The unit must be kept out of areas that are subject to freezing temperatures 2 High temperatures greater than 100 F 38 C must be avoided If the unit is used outside avoid puttin...

Страница 4: ...Low water temperature Heat feed water or use higher GPD membrane Low feed pressure Use booster pump or use higher GPD membrane Fouled membrane Replace membrane Membrane Troubleshooting Guide The follo...

Страница 5: ...3 X Y 143 ppm 4 X Y X 143 150 0 953 5 Rejection X Y X 100 0 953 100 95 3 Membrane Hardness Rejection 95 3 Membrane OK Production Rate SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 87...

Страница 6: ...illiliters per Minute 5 If the actual concentrate flow rate is less than the optimum concentrate flow rate you must reduce the length of the flow restrictor capillary tubing as described on the next p...

Страница 7: ...ncentrate tubing Fig C for Single Membrane System or Fig D for Dual Membrane System Remove the tubing from its push fitting at the membrane as follows a Firmly depress and hold the push fitting collar...

Страница 8: ...ail and remove the blue product water inter connect tubing 9 Lift the dual RO membrane assembly from the holding clips 10 Separate the two membranes by depressing the collar on the push fitting of the...

Страница 9: ...mbrane 1 Membrane 2 feed push fitting black inter connect feed tubing blue product water interconnect tubing Membrane 1 concentrate push fitting Membrane 1 product water push fitting SpectraPure Spect...

Страница 10: ...raPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 15 Note The flow restrictor assembly should remain uninst...

Страница 11: ...as described in the Removal Adjustment and Replacement section on page 7 Snap the membrane assembly back into its holding clips System Maintenance RO Membrane Replacement If you have a Single Membran...

Страница 12: ...and check for leaks System Maintenance SpectraPure SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 13 System Mai...

Отзывы: