MEGA MAXCAP DI REPLACEMENT:
The Mega MaxCap™ DI cartridge is located in the second housing from the right.
When the Right-Side Meter (set to the “IN” position) reads “1”, it is time to replace
the Mega MaxCap™ DI cartridge.
1. Turn the system off and remove the cartridge housing by rotating it in a
counterclockwise direction.
2. Thoroughly wash out the housing with hot soapy water to which a few
teaspoons of household bleach have been added. Rinse well with clean
hot water.
3. Make sure the DI cartridge is installed in the correct direction. The white cap
should be facing upward. Make sure that the top seal is securely attached
to the top of the cartridge.
4. Tighten the cartridge housing by rotating it clockwise and hand tighten.
5. Turn on system and check for leaks.
ENDURO DI REPLACEMENT:
The Enduro DI™ cartridge is located in the rightmost housing. When the Right-
Side Meter (set to the “OUT” position) reads “1”, it is time to replace the Enduro
DI™ cartridge.
1. Turn the system off and remove the cartridge housing by rotating it in a
counterclockwise direction.
2. Thoroughly wash out the housing with hot soapy water to which a few
teaspoons of household bleach have been added. Rinse well with clean
hot water.
3. Make sure the DI cartridge is installed in the correct direction. The white cap
should be facing upward. Make sure that the top seal is securely attached
to the top of the cartridge.
4. Tighten the cartridge housing by rotating it clockwise and hand tighten.
5. Turn on system and check for leaks.
Mega MaxCap DI
Top Seal
Enduro DI
Right-Side
TDS Meter
Enduro DI cartridge
shown with white cap
facing upward