MAXCAP SUPER DI MAINTENANCE AND REPLACEMENT:
1. When the reading on the right-hand TDS meter (set to “IN”) displays 50% of
the reading on the left-hand TDS meter (set to “OUT”), it is time to replace
the MaxCap
®
Extended Life Super DI cartridge (DI-MC-10HC).
2. Turn the system off and remove the cartridge housing by rotating it counter-
clockwise until it is free of the threaded head on the bracket.
3. Make sure the DI cartridge is installed in the correct direction as marked on
the cartridge shell and be sure that the top seal is securely attached to the
top of the cartridge.
4. Tighten the cartridge housing by rotating it clockwise and hand tighten.
5. Turn on system and check for leaks.
SILICABUSTER SUPER DI MAINTENANCE AND REPLACEMENT:
1. When the reading on the right-hand TDS meter (set to “OUT”) displays
“001”, it is time to replace the SilicaBuster™ Extended Life Super DI cartridge
(DI-SB-10HC).
2. Turn the system off and remove the cartridge housing by rotating it counter-
clockwise until it is free of the threaded head on the bracket.
3. Make sure the DI cartridge is installed in the correct direction as marked on
the cartridge shell and be sure that the top seal is securely attached to the
top of the cartridge.
4. Tighten the cartridge housing by rotating it clockwise and hand tighten.
5. Turn on system and check for leaks.