8194B, 8195B, 8197B
Spectracom Corporation
GPS Ageless Master Oscillator Instruction Manual
4-30
4.5.3 Alarm
Status
This command read
Alarm Status
. This is a short command designed to let the
controller know if there are any alarms present.
rast
rastC
where C = y or n indicating if there are alarm conditions present.
4.5.4 Alarm
Timeouts
This command is to read or set the three
Alarm Time outs
individually. Each time-out
controls the status of an alarm relay. If the GPS receiver stops tracking satellites the
timers start counting down. If the timer reaches 0 before the receiver begins to track
satellites again, the alarm relay is activated. In addition, if Alarm Time out 1 occurs, this
is a minor alarm. Alarm Time out 2 and Alarm Time out 3 are Major alarms. Defaults
are: AT1 = 1 minute, AT2 = 2½ hours, AT3 = 30 days.
watXDDDHHMMSS
where X = 1,2, or 3 indicating which time out.
watXDDDHHMMSS
ratX
ratXDDDHHMMSS
4.5.5 Clear Alarm History
This command is to
Clear the Alarm History
. This command accompanies the
raeh
command and is used to “erase” the alarm history. After this is issued, subsequent use
of the
raeh
command will respond with the valid entry character set to ‘n’, until a new
alarm event occurs.
wcah
wcah