8194B, 8195B, 8197B
Spectracom Corporation
GPS Ageless Master Oscillator Instruction Manual
3-22
1
15V
230V
Figure 3-17 AC Power Module
3.3.12 Cooling fan
Cooling fan operation is microprocessor controlled. The fan is turned on when the
internal temperature measures 50°C and off when the temperature reaches 30°C.
3.3.13 Chassis Ground
The chassis ground lug allows the unit’s chassis to be connected to an earth ground in
addition to the power line safety ground. Connecting the chassis to a single point
ground system may be required in some installations to ensure optimum lightning
protection. A separate earth ground is also recommended in installations where
excessive noise on the power line degrades the unit’s receiver performance.
3.3.14 Battery Disconnect Switch
Depressing this switch removes the backup battery power from the GPS receiver and
ovenized oscillator. The Battery Disconnect switch is operational only on units equipped
with Option 02 and when AC power is removed. This feature prevents battery
discharge when shipping or storing the unit.
NOTE:
The internal battery must be disabled prior to shipment or storage. Failing to
disable the battery may result in reduced capacity and shortened battery life.
NOTE:
The Battery Disconnect Switch is populated on all versions of the Model 8195B.
The switch is only operational on units equipped with Option 02.