Cirrus SP101013.111 – 10 October 2012
MKS Instruments UK Ltd., Spectra Products
22
5.3 Re-setting the IP Address
Note:
This is not an operator function, only competent persons may carry out this operation
Re-setting the IP address back to the default, will require the removal of the Cirrus cover as detailed in
Section 7.1.
Once the cover has been removed, locate the “preset” and “reset” buttons located on the top left hand
side of the main PCB.
With power applied to the Cirrus, press the “reset” button and immediately press and hold the “preset”
button until a “beep” is heard from the unit.
Press the “reset” button to reboot the Cirrus which will now use its default IP address.
Note:
If the unit has been configured to make use of DHCP, this procedure can take up to
90
seconds
before the beep is heard.
Replace the cover before allowing operator access.