SVC HR-1024i / HR-768i / HR-640i / HR-768si / HR-512i USER MANUAL REV 1.12
S P E C T R O R A D I O M E T E R
4 4
4 4
4 4
4 4
Note that this button will
only
appear if
a) the wave cal file was just read in from
disk, and b) the wave cal file has the
correct corresponding “key” file
associated with the “cal” file. Contact
Spectra Vista for further information.
Radiometric Calibration Tab
:
There are up to 8 radiometric calibration
tables (corresponding to up to 8 optional
instrument fore-optics) stored within the
instrument.
This tab displays each of these eight
1024-entry tables, one table at a time. The
table that is currently displayed may be
selected via the “
Select Optic Number
”
control at the top of the tab.
Note that
buttons that are used to read or write
radiometric data only operate on a
single table at a tim
e – the table
currently selected by the “Select Optic
Number” control.
Each fore-optic’s radiance calibration
information has an “
Optic Name
”,
which is a 6-character familiar name that
is displayed on the instrument’s LCD.
Each fore-optic calibration also has an
“
Optic Type
” – the radiometric
calibration data is either a “Radiance”
Содержание SVC HR-1024i