equipment damage. 

13.

 

Do not use harsh cleansers, solvents, or detergents.  Do not expose the unit to excessive moisture. 
Equipment damage could occur. 

 
 

Warning: This product contains/may contain chemicals known to the state of California to cause cancer 

and/or birth defects or other reproductive harm. 

 
 

 

We believe the PressureGuard Easy Air sets a new standard for 

low air loss mattress systems. 

 

Thank you for choosing Span-America! 

 

 

Содержание L7535-29

Страница 1: ...09564 R4 CO 2340 OWNER S MANUAL PressureGuard EASY AIR Series Span America Medical Systems Inc 800 888 6752 www spanamerica com PressureGuard EASY AIR PressureGuard EASY AIR LR PressureGuard EASY AIR XL ...

Страница 2: ...2 P09564 R4 CO 2340 ...

Страница 3: ...3 DIRECTIONS FOR USE EASY AIR STANDARD 14 EASY AIR XL 14 EASY AIR LR 16 GENERAL DIRECTIONS FOR USE ELEVATING HEAD OF BED 17 BED RAILS 17 CPR 17 PATIENT TRANSPORT 17 POWER LOSS 17 BED LINENS 17 INCONTINENCE PADS 17 SERVICE 17 ENVIRONMENTAL CONDITIONS FOR USE 17 STORAGE AND TRANSPORTATION 18 CLEANING MATTRESS COVERS 18 AIR CONTROL UNIT 19 PREVENTIVE MAINTENANCE FILTER MAINTENANCE 20 POWER CORDS 21 M...

Страница 4: ...4 P09564 R4 CO 2340 ...

Страница 5: ...uction clarification The symbol below identifies a WARNING or CAUTION A WARNING identifies situations or actions that may have an effect on patient or user safety Ignoring a warning could cause patient or user injury A CAUTION points out special procedures or precautions that persons must obey to avoid equipment damage The symbol below identifies an ELECTRICAL SHOCK HAZARD WARNING The symbol below...

Страница 6: ...ss can be turned ON or OFF Standard Easy Air and Easy Air XL bariatric versions provide choice of two therapy modes of operation FLOAT or ALTERNATING Easy Air LR version provides choice of three therapy modes of operation FLOAT ALTERNATING or ROTATION To reduce the risk of burns electrocution fire or injury to persons READ ALL INSTRUCTIONS BEFORE USING THIS UNIT 1 Use this unit only for its intend...

Страница 7: ...essive moisture Equipment damage could occur Warning This product contains may contain chemicals known to the state of California to cause cancer and or birth defects or other reproductive harm We believe the PressureGuard Easy Air sets a new standard for low air loss mattress systems Thank you for choosing Span America ...

Страница 8: ...he surface to the mattress maintenance is minimal The air cylinder inflation system and the foam shell work in concert to maintain low interface pressures throughout the surface making the Easy Air effective for treatment of pressure injuries by preventing further tissue breakdown Close up illustrations of the Outer Removable Coverlet and the Inner Air Delivery Cover and an explanation of the Scie...

Страница 9: ...r cylinders that run lengthwise underneath the body Unlike typical low air loss systems the cylinder inflation system does not lose air from the inflation system for low air loss and can be programmed to perform either Flotation or Alternating Pressure in the LR model they can also perform Lateral Rotation The cycles and inflation levels are designed to provide and maintain low interface pressures...

Страница 10: ...through the cover where a continuous air current takes it away before it can re form as liquid The Air Diffusion Matrix fabric is not collapsible ensuring a pathway for a constant flow of air beneath the patient Compare to typical low air loss designs that cause the patient s body to press the cover directly onto the air holes closing off the flow of air beneath the patient Outer Removable Coverle...

Страница 11: ...facility engineering department Multiple outlet strips should be mounted on a fixed object to reduce the risk of liquid spills and physical damage In addition if multiple receptacle outlet boxes are used they also should be protected from the risk of liquid spills and physical damage All extension cords and multiple outlet strips should be tagged and inspected routinely Do not cover the power cord...

Страница 12: ...ing of the air lines or the motion of the bed does not impede air flow by crimping the air lines ELECTROMAGNETIC OR OTHER INTERFERENCE This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to other devices in the vicinity However there is no guarantee that interference will not occur in ...

Страница 13: ... firmly for 3 4 seconds until light turns off indicating that the control unit settings are again unlocked and can be adjusted as desired 5 COMFORT LEVEL For best possible pressure management maximize immersion and envelopment by initially setting the COMFORT LEVEL on the control panel to the softest selection appropriate for the patient in accordance with the stated weight limits Easy Air Easy Ai...

Страница 14: ...r THERAPY MODE on the control unit s front panel Select ALTERNATING Air cylinders 1 and 3 deflate while 2 and 4 inflate Pairs will go through one complete alternation cycle every 15 minutes OR FLOAT evenly inflates all 4 air cylinders providing powered flotation 3 Press the AIR FLOW toggle button to select AIR FLOW ON or OFF as preferred AIR FLOW is ON when light is illuminated and OFF when light ...

Страница 15: ...ill go through one complete deflation inflation cycle every 22 minutes OR FLOAT evenly inflates all 6 air cylinders providing powered flotation 3 Press the AIR FLOW toggle button to select AIR FLOW ON or OFF as preferred AIR FLOW is ON when light is illuminated and OFF when light is not illuminated 4 In normal operation the green NORMAL PRESSURE indicator should be illuminated If the amber LOW PRE...

Страница 16: ...the alternating or rotating movement of the mattress and evenly inflates all four air cylinders The FLOAT indicator light will illuminate and the mattress will achieve uniform support with no movement within five minutes ROTATION Air cylinders 1 and 2 inflate while 3 and 4 deflate gently rolling the patient laterally Cycles complete every 15 minutes rotating the patient from side to side approxima...

Страница 17: ...ulmonary Resuscitation The face plate of all Easy Air models indicates recommended settings for CPR The standard for life support recommended by the American Heart Association recommends a hard level surface such as a crashboard or moving the person to the floor if possible For performing CPR on all PressureGuard Easy Air Models 1 Select the FLOAT Mode 2 Select the firmest COMFORT LEVEL 3 Place a ...

Страница 18: ...term incontinent applications clean and disinfect cover daily A scented cleaner disinfectant is recommended Iodophor type disinfectants e g Betadine will stain the fabric For disinfection phenolic or quaternary type disinfectants are recommended Disinfectants should be hospital grade tuberculocidal Follow manufacturer s instructions for use concentrations contact times and rinsing Contamination wi...

Страница 19: ...setting when powered back up Wipe down with using damp sponge or cloth that has been thoroughly wrung out to remove excess liquid Do not allow liquids to penetrate the user panel For cleaning use neutral suds and lukewarm water For disinfection phenolic or quaternary type disinfectants are recommended Disinfectants should be hospital grade tuberculocidal Follow manufacturer s instructions for conc...

Страница 20: ...and to prolong the life of the unit cleaning of the two part filter system should be performed at least once every 180 days of use If the control unit is used in an area with high concentration of dust or cigarette smoke the filter should be cleaned as frequently as once every 30 or 60 days for maximum operation TO CLEAN THE TWO PART FILTER SYSTEM 1 Turn unit OFF Remove the four screws on the filt...

Страница 21: ...used and that they should be mounted with the ground pin or neutral blade up to assure that any metal that may drop between the plug and the wall will most likely contact an unenergized blade 3 Check the power cord to assure that contact pins are straight and secure 4 Routinely inspect the power cord for damage sustained from crushing pinching shearing cutting or from being worn through They can b...

Страница 22: ... or torn inspect the internal components for signs of contamination If contamination is evident quarantine the mattress and remove from service following infection control procedures You may use the Preventive Maintenance Log provided on the last page of this manual to monitor and document regular inspection and maintenance of your PressureGuard Easy Air Systems ...

Страница 23: ...vides pressure management in the prevention and treatment of pressure injuries The surface is designed to prevent bottoming out for patients weighing up to 500 lbs The low air loss feature reduces excessive perspiration and body heat by transporting moisture vapors away from the patient s skin OPERATING MODES Air Flow Select from OFF or ON all models Therapy Mode Select from FLOAT or ALTERNATE all...

Страница 24: ...ress Size Mattress Only Removable Coverlet Only Inner Cover Replacement Inflation System L7535LR 29 75 L x 35 W x 7 H LM7535LR 29 CLT L7535LR C1 L7535 29 P09275 L8035LR 29 80 L x 35 W x 7 H LM8035LR 29 CLT L8035LR C1 L8035 29 P09138 L8435LR 29 84 L x 35 W x 7 H LM8435LR 29 CLT L8435LR C1 L8435 29 P09138 L8039LR 29 80 L x 39 W x 7 H LM8039LR 29 CLT L8039LR C1 L8039 29 P09138 L8053LR 29 80 L x 53 W ...

Страница 25: ...uires 2 each P02813 Small Female while plastic 8200 model requires 2 each 8300 model requires 4 each P02337 Power cord all models requires 1 each P09532 Power Module Mattress End P02404 Small Male Elbow 8000 model requires 2 each not shown P02374 Large Male all models requires 1 each P09312 Small Male metal 8200 and 8300 models require 2 each P02800 Small Male white plastic 8200 model requires 2 e...

Страница 26: ...e knees Adjust each as necessary to meet patient needs while maximizing turn angle Patient exceeds weight limit Call Span America for assistance with product selection Mattress not inflating or patient reports a sinking feeling Control Unit is not turned on Turn control unit on Airlines not connected Ensure secure connection of airlines at control unit and mattress Airlines or quick disconnect con...

Страница 27: ... than one button on overlay Reset control unit by turning off then back on Call for service if problem not resolved Central processor malfunction Reset control unit by turning off than back on Call for service if problem not resolved Interference produced to electronic equipment devices in surrounding area Electromagnetic interference caused by the unintentional emission of electromagnetic waves o...

Страница 28: ...28 P09564 R4 CO 2340 ...

Страница 29: ...09564 R4 CO 2340 PressureGuard Easy Air Preventive Maintenance and Repair Log Date Air Filter Power Cord Mattress Repair Manufacturer Span America Date Purchased Serial C Cleaned OK Okay R Repaired Replaced ...

Страница 30: ...30 P09564 R4 CO 2340 ...

Страница 31: ...31 P09564 R4 CO 2340 ...

Страница 32: ...32 P09564 R4 CO 2340 THIS PRODUCT IS PROTECTED BY ONE OR MORE OF THE FOLLOWING U S PATENTS 6 782 574 6 223 369 6 848 138 7 296 315 SPAN AMERICA MEDICAL SYSTEMS INC Greenville SC 29615 800 888 6752 ...