PREVENTIVE MAINTENANCE
FILTER MAINTENANCE
• The control unit contains a high volume air blower that requires filtered air for proper
operation.
• The two-part filter system is located on the bottom rear of the unit.
•
The larger outer filter is designed to be removed, cleaned, and re-used. [Fig. 1 & 2]
•
The smaller inner filter is located directly beneath the larger removable filter [Figure 3]. THE SMALLER
FILTER IS NOT TO BE REMOVED. It is designed to be cleaned in place, from outside of the unit.
CAUTION: For maximum effectiveness, and to prolong the life of the unit, cleaning of the two-part filter
system should be performed at least once every 180 days of use. If the control unit is used in an area with
high concentration of dust or cigarette smoke, the filter should be cleaned as frequently as once every 30 or
60 days for maximum operation.
TO CLEAN THE TWO-PART FILTER SYSTEM
1. Turn unit
OFF
. Remove the four screws on the filter housing plate, which is found at the bottom rear of the
control unit [Figure 1].
Figure 1
Figure 2
2. With plate off, pull filter out [Figure 2].
3. Wash filter using mild soap and water. Pat dry with towel.
NOTE
: If filter is damaged or cannot be easily cleaned, it should be discarded and replaced with a new filter,
Span-America part number P02549.