![SPAN L7535-29 Скачать руководство пользователя страница 15](http://html.mh-extra.com/html/span/l7535-29/l7535-29_owners-manual_1339367015.webp)
15
P09564 R4 CO# 2340
DIRECTIONS FOR USE
PRESSUREGUARD EASY AIR
®
XL MODEL
1. Immediately after the control unit is switched
ON
, all lights on the control panel will illuminate for 1-2
seconds while the system performs a self check. If correct system air pressure is not reached within 60
seconds, the amber
LOW PRESSURE
indicator will illuminate. LOW PRESSURE indicator will remain until
being replaced by the green NORMAL PRESSURE indicator, indicating that the mattress is ready for patient
use. If NORMAL PRESSURE indicator does not illuminate within 20 minutes of power up, contact Span-
America for assistance.
2. Select the preferred mode of operation by pressing the toggle button under
THERAPY MODE
on the control
unit’s front panel. Select:
ALTERNATING
: The six cylinders work in pairs. Air cylinders 1 and 4 deflate, then 2 and 5, then 3 and 6.
Pairs will go through one complete deflation/inflation cycle every 22 minutes.
OR
FLOAT
evenly inflates all 6 air cylinders, providing powered flotation.
3. Press the
AIR FLOW
toggle button to select
AIR FLOW ON
or
OFF
as preferred.
AIR FLOW
is ON when light
is illuminated, and OFF when light is not illuminated.
4.
In normal operation, the green
NORMAL PRESSURE
indicator should be illuminated. If the amber
LOW
PRESSURE
indicator illuminates, verify air lines are not crimped or kinked preventing proper air flow and
that all air line connections are secure. If the problem is not resolved, remove the patient from the surface
and discontinue use. For assistance, contact Span-America at 800-888-6752 8:00 AM – 5:00 PM EST Monday
through Friday.
5.
On the XL model, the
COMFORT LEVEL
is not intended to be set according to the individual user’s specific
weight. Instead, for best possible immersion and pressure distribution, begin with the softest setting, setting
1. If user expresses discomfort, level can be adjusted as desired to accommodate comfort.