Appendix A — Electromagnetic Compatibility
91518 Multigas Analyzer Service Manual
A-2
Electromagnetic Immunity
Note:
This equipment is intended for use in the electromagnetic environment specified in the table below. The
customer or user of the equipment should ensure that it is used in such an environment
.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/output
lines
±1 kV for input/output
lines and patient
cables
Mains power quality should be
that of a typical commercial or
hospital environment.
Refer to the host
monitor/module housing service
manual.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common mode
±1 kV differential
mode
±2 kV common mode
Voltage dips,
short
interruptions,
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
for
0.5 cycle)
40% U
T
(60% dip in U
T
for
5 cycles)
70% U
T
(30% dip in U
T
for
25 cycles)
<5% U
T
(>95% dip in U
T
for
5 seconds)
<5% U
T
(>95% dip in U
T
for
0.5 cycle)
40% U
T
(60% dip in U
T
for
5 cycles)
70% U
T
(30% dip in U
T
for
25 cycles)
<5% U
T
(>95% dip in U
T
for
5 seconds)
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
60 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a commercial or
hospital environment.
Note
:
U
T
is the AC mains voltage prior to application of the test level. All power line immunity tests
were performed on the host monitor/module housing at 120 VAC/60 Hz and 230 VAC/50 Hz.