91518 Multigas Analyzer Service Manual
5-3
Troubleshooting
SAMPLING
SUSPENDED
Patient is not
monitored
SERVICE REQUIRED -
Send for repair
■
Defective hardware.
■
Multigas analyzer self-test
failure.
■
Invalid option programming.
■
Invalid or corrupt EEPROM
configurations
■
Check for power to the multigas
analyzer and the interface board
as follows:
•
J501:
Pin 7-12 V ±0.50 V
Pin 10+12 V ±0.50 V
Pin 11+5 V ±0.25 V
•
J720
:
Pin 5-12 V ±0.50 V
Pin 7+3.3 V ±0.25 V
Pin 8+12 V ±0.25 V
Pin 9+5 V ±0.25 V
■
Contact a Spacelabs Healthcare
field service engineer
COMMUNICATION
LINK LOST - Service
required
-or-
COMMUNICATION
DROP OUTS - Service
required
■
An internal communication
error was detected.
■
Troubleshoot multigas analyzer;
start by replacing OEM multigas
bench.
■
Return the multigas analyzer to
Spacelabs Healthcare for repair.
SERVICE CAL MODE -
Patient is not
monitored
■
Service Calibration
menu is
displayed
■
Touch NORMAL SCREEN to exit
this mode.
NO BREATH
DETECTED
■
Breaths are not detected for
a duration equal to the
APNEA timeout duration
■
Verify sensor is properly
connected to the patient.
Message will clear after three
breaths are detected, or the
ACKNOWLEDGE ALARM key is
touched.
CAL GAS IN
PROGRESS
■
CAL GAS calibration is in
progress.
■
Wait until CAL GAS calibration is
completed and the message
goes away.
ZERO IN PROGRESS
■
A Zero calibration is in
progress.
■
Wait until Zero calibration is
completed and the message
goes away.
Clinical Situation
Possible Cause
Solution