Maintenance
91518 Multigas Analyzer Service Manual
4-10
Note:
There is a two-position dip switch located on the board edge near the end of the CPU board used with the
EEPROMs. It is shown in the normal operating position in Figure 4-11. Be careful during preventive
maintenance to ensure that the switch positions are not changed.
Figure 4-11: Dip switches, normal operating position
Pneumatic Block Filter Replacement
To replace the pneumatic block filter:
1 Disassemble the gas unit according to
on page 4-7. Place the gas unit on an
anti-static mat with the pneumatic block and nafion tubing side facing up.
2 Remove the small Phillips head screw holding the filter restraint (refer to
). Use a needle nosed
pliers or tweezers to lift the filter restraint away from the block.