Chapter
2
Revision 001, 03/09
3
© SP Industries, Inc. 2009
Navigation
The Wizard 2.0 controller is divided into four primary areas.
•
The Manual Mode buttons allow you to activate various components of your
freeze dryer.
•
The LCD display shows the status of the freeze dryer in real time and
indicates recipe progress and any active alarms.
•
The tactile buttons below the LCD display allow you to enter recipe and
configuration data, as well as navigate system menus.
•
The Automatic Mode buttons allow you to control your freeze dryer, as well as
navigate through various menus and screens of your Wizard 2.0 controller.
Button Functions
Manual Mode Buttons
•
Freeze
activates all components needed to cool the shelves to the Manual
Freeze setpoint.
•
Condenser
activates the condenser refrigeration system and cools the
condenser.
•
Vacuum
activates the vacuum system. Use in conjunction with the
Vac Ctrl
button to control the vacuum pressure at a defined setpoint.
Note:
Select your preferred unit of measure—millitorr, microbar or Pascal—from the
Configuration menu
•
Shelf Ctrl
initiates shelf temperature control at the Manual Heat setpoint.
Note:
The
Shelf Ctrl
button should be used with the
Condenser
button
to control the
shelf temperature at the setpoint.