VirTis Wizard 2.0
Appendix B: Factory Settings
Revision 001, 03/09
27
© SP Industries, Inc. 2009
PID Control Settings
These settings are used in the proportional-integral-derivative (PID) tuning
calculations for the heating and cooling controls.
Suggested settings are as follows:
Model
P
I
D
AdVantage 2.0
5
16
1
AdVantage Plus
2
16
0
Genesis, Ultra and VirTual
2
16
1
Note:
PID Control Settings are factory-set based on performance during quality control testing.
Your settings may not match the settings above. SP Industries recommends recording your
factory default settings before making any changes.
Relay Test
The Relay Test menu displays the four output groups. Use this screen for
troubleshooting electrical or mechanical problems, as well as during a freeze-drying
run to observe active outputs.
SSR 1
SSR 2
SSR 3
SSR 4
Vacuum
Heater Defrost Alarm
Output
Vacuum Release
Condenser Valve Fluid
Pump
Stoppering
^
Compressor 1
Compressor 2
Spare
Stoppering -
Freeze Valve
Negative Shelf Isolation
Valve Spare
Pressing the corresponding number on the numerical keypad will begin a test of each
SSR group. The outputs in this SSR group will energize in sequence for
approximately two seconds each. A dark square appears next to an active output.
Notes:
All control loops are disabled unless you are viewing the Synoptic screen.
Only a qualified SP Industries Technician should interpret the results of a relay function test.
Not all functions are available on all units.
Содержание VirTis Wizard 2.0
Страница 1: ...SP INDUSTRIES INC Controller Operator s Manual VirTis Wizard 2 0 Lyophilizer Controller ...
Страница 2: ......
Страница 6: ...iv Revision 001 03 09 SP Industries Inc 2009 ...
Страница 36: ...SP Industries 815 State Route 208 Gardiner NY 12525 USA www SPIndustries com 800 431 8232 845 255 5000 ...