ST2883 Series Operation Manual Chapter 5 Introduction to [SETUP]
5-4
5.2.8 Delay Time
Delay time refers to the time between the end of a measurement and the start of the next measurement. The
delay function is only available when trigger mode is set to INT trigger mode. The trigger delay time can be
set from 0.1s to 99s by 0.1 steps.
Move the cursor to this zone. User can input the delay time by pressing numerical keys and ENTER for
confirmation. The input unit is s. Following soft keys can also be used to set the delay time.
They are coarse adjustment keys used to change the delay time by 1.0s.
They are fine adjustment keys used to change the delay time by 0.1s.
Lighten this key to enable the numerical arrow key. Respectively press 8, 2, 4, 6, the cursor will move
up, down, left and right.
5.2.9 BDV Test
BDV test is a test method for turn-to-turn insulation test. Starting from Start Volt and ending at End Volt by a
resolution of Volt Step, the test is finally done by comparing with four comparison methods. Move the cursor
to this zone, following soft keys will be displayed.
ON Turn BDV test on.
OFF Turn BDV test off.
Lighten this key to enable the numerical arrow key. Respectively press 8, 2, 4, 6, the cursor will
move up, down, left and right.
Note: Only in ON mode, Start Volt, End Volt and Volt Step can be set.
5.2.10 Start Volt
Start Volt ranges from 100V to 5000V but cannot be larger than End Volt. Move the cursor to this zone,
following soft keys will be displayed.
They are coarse adjustment keys used to change Start Volt by 100V.
They are fine adjustment keys used change Start Volt by 10V.
Lighten this key to enable the numerical arrow key. Respectively press 8, 2, 4, 6, the cursor will move
up, down, left and right.
5.2.11 End Volt
End Volt ranges from 100V to 5000V but must be larger than Start Volt. Move the cursor to this zone,
following soft keys will be displayed.
They are coarse adjustment keys used to change End Volt by 100V.
They are fine adjustment keys used change End Volt by 10V.
Lighten this key to enable the numerical arrow key. Respectively press 8, 2, 4, 6, the cursor will
move up, down, left and right.