14
7. RS232 output
The equipment has an RS232
data output in the control panel that can
be used to register measurements in a serial printer or data collector. The
outputs are the following ones:
Pin 2: Rx ; Pin 3: Tx ; Pin 5: Gnd;
Rate: 4800 bps
8 bits - no parity - 1 stop bit (8,n,1)
Note: In order to assure the compatibility with most printers available in
the market, the resistance units are shown with the following symbols:
uR = micro-ohm
mR = milli-ohm
R = ohm
8. Cleaning
The panel, terminals and connectors of the equipment must stay dry and
clean. Cleaning should be made using a wet cloth in water and a soft
detergent or isopropyl alcohol (be sure that the products to be used for
cleaning does not affect plastic goods).
9.
Replacement fuse
To check the instrument fuse
, remove it with a screw driver. If the fuse
is ruptured replace it by another with the following specifications:
Fuse Schurter, model SPT 5x20 (Time-lag) 5A/250V. High breaking capacity.
Содержание MPK-2000e
Страница 16: ...18 Notes...