11
6. Turning the test current control
clockwise increases the current.
Adjust this control until obtaining the desired value or until the current
indicator (bargraph) indicates 100%. The bargraph will show the test
current value as a percentage of the nominal value selected by using
the test current selector
.
7. The lowest current for measuring is 10% of the nominal value. It is
important to consider that the measurement errors increase while test
current decreases. The equipment accuracy is specified for test
currents higher than 80%.
8. The display
will show the resistance value measured and the
corresponding unit (
Ω
[ohms], m
Ω
[milli-ohms] or µ
Ω
[micro-ohms]).
9. The value can be retained in the display by pressing the hold
key.
Pressing this key again, the value will be released.
10. Press the Stop
key in order to finish the measurement. Do not turn
Off the equipment without pressing the Stop
key before.
11. Finally, when finishing measurements, turn the equipment Off using
the On/Off
switch.
Precaution: Do not connect or disconnect the test leads during the
measurement.
Содержание MPK-2000e
Страница 16: ...18 Notes...