5. Messages
SOURCETRONIC MPK-2000e
When turning the equipment On using the On/Off
switch, this
introductory message appears for a while. During that time, the
equipment carries out some functional checking.
WAIT...
This message appears each time the equipment needs to adjust any
parameter in order to optimize the readings.
PRESS START
The equipment is ready to start a measurement, thus the operator has to
press the Start
key.
LOW CURRENT
It shows that the test current is not enough to carry out the reading. It
appears at the beginning of each test and it keeps on being there up to
the operator rotates the test current control
clockwise, as necessary
for the test current to be higher than 10 % of the nominal current in the
scale. The inappropriate connection of the cables may cause a difficult
circulation of test current. If this message keeps on being displayed,
please check that the current cables are connected appropriately.
OVERRANGE
It indicates that the measured resistance is higher than the maximum
value readable in the selected range.
H
Indicates that the value is the one retained in the memory when pressing
the hold
key.
BAT
It shows that the battery is quite discharged. It is necessary the battery to
be recharged.