18
3.12. Auto power-off
The Auto-Power-off function turns off the equipment consumption
(independently of the timer function) in two situations:
During the measurement
- After 35 minutes of operation, without
allowing that the battery status checking function is carried out during
that period.
Idle equipment
- After 10 minutes of inactivity.
4. Battery charger
This equipment has an intelligent built-in circuit that controls the battery
charge and doesn’t allow the equipment to operate during the charging
process. In order to charge the battery, follow this procedure:
Verify that the
On/Off
switch is switched off.
Connect the power supply cable plug into
input
. After a while, the
battery charge indicator ( key led)
will blink alternatively in green and
red during one second, while the charger verifies the initial condition of
the battery to select the optimised parameters of the charge.
The following chart summarizes the meaning of LED luminous indications:
Green and red
flashing alternatively
Test of the initial condition of the battery when plugging
the mains, during one second.
Permanent red
Battery under charge.
Flashing red
Charging current is less than normal.
Содержание MD-5060 Series
Страница 24: ...26 8 Printer Printing sample...