GasLab Q2 displays and operating the device
Page 90 of
135
G a s Q u a l i t y A n a l y z e r
GasLab Q2
Identifier
Root Cause of Q2 AFB messages
Alarm Master-Slave
parameterization error
An error event has occurred the alarm boundaries (limits) are
violated. No valid measurement result is displayed. The final
value
can’t be calculated correctly.
Rectification
Transfer a valid parameterization into the device and reboot it. If
you do not succeed in this way, please contact Elster GmbH
Other Master-Slave Alarms
As a rule, it typically involves subsequent errors resulting from
other causes. Please contact Elster GmbH
Alarm
Q2SENS communication timeout
Alarm Q2SENS write error
Q2SENS link lost
Q2BASE link lost
Alarm signal
Q2BASE write error
The adjacent alarms indicate communication and memory errors.
Rectification
Interrupting the power supply and reboot, repair by Elster service
generally required. Please contact Elster GmbH
Identifier
Root Cause of Q2 AFB messages
Warning
Standard A failure
The final value
can’t be calculated correctly.
Rectification
Please contact Elster GmbH.
Warning
Standard B failure
Internal warning
The Basic System has a fault. The device does not operate
smoothly.
Rectification
Please contact Elster GmbH
Internal alert
Identifier
Root Cause of Archive AFB message
The User archeve system has a fault. Results could not saved
and archived in a safety maner.
Rectification
Please contact Elster GmbH
Identifier
Root Cause of Modbus AFB message
Protocol error
The protocol has an error
Rectification
The Hint disappears automatically.
The user can also generate further display texts for each type. It is done in
Postprocessing AFB by changing the parameterization. For more information
about parameterization and Postprocessing AFB
see chapter 8