Transport, storage, installation
5
GasLab Q2
G a s Q u a l i t y A n a l y z e r
Page 31 of 135
No. Gas Connection
1
Process gas (PG) inlet
Inlet pressure: 0.15 to 0.5 MPa
(gauge)
with an internal sintered particle
filter
2
Calibration gas in cylinder (CAL)
Operational calibration gas
(CH4 + CO2)
0.2 MPa (g) ±0.05 MPa
Maximum inlet pressure:
0.25 MPa (gauge)
3
Integrated bypass outlet
Flow: up to approx. 300 l/h
(if necessary)
4
Exhaust gas outlet
Standard measurement path
sensor
Flow: typical 30 l/h
Flow: max. 60 l/h
5
Main breather
with thread: ¼” NPT
for piping, min. ID 8 mm, Length
according to chapter 2.4
If piping is connected, this must be
separate from other exhaust piping
6
Rear breather
Integrated in rear side of
connection box
No piping possible nor required
Figure 5.2: Definition of gas connections
See chapter 4.2.1 and 10.
Содержание Multiscan CPD-1304S
Страница 1: ...73023639 g 2017 03 GasLab Q2 Ex d ...