GasLab Q2 displays and operating the device
7
GasLab Q2
G a s Q u a l i t y A n a l y z e r
Page 67 of 135
In the top line on light blue background, the operating mode is indicated next to "
Operation
". Four
“
Operations
” are usually used by the operator. Beside "
Step
" the actual status of device working
step is shown too. Seven "
Steps
”are usually used. Following combinations are possible:
Operation
Step
Meaning
Startup
Startup
Device in boot sequence
Analysis
Flushing
Cleaning of the gas paths with gas for next measurement
Analysis
Normal Operation (result finding)
Operational
Calibration
Start
Wait for confirmation before beginning
Flushing
Cleaning of the gas paths with gas for next measurement
Gas 1
Result finding / calibration
End
Wait for confirmation after finish or interruption of operation
Verification
Start
Wait for confirmation before beginning
Flushing
Cleaning of the gas paths with gas for next measurement
Verification
Result finding / verification
End
Wait for confirmation after finish or interruption of operation
More details about the listed
“Operations”and “Steps” are given, when
used, later on in the text.
The device still has other operating modes and steps which are not
specified and described here, since they are only used by Elster GmbH.
It is possible to display a second calculation standard and switch between these standards, so
the appearance of the correspondending line is variable, see the following examples with two
standards.
The current standard with its settings is
shown middle, in the example ISO 6976
(Standard A). On the right hand side a link
to the alternative standard B is shown. It can
be selected. The title (Main display)
indicates additional, that the results of the
standard A are shown in the adjacent
picture.
Activate
Standard B
. Use the overlay keys. By this changeover following changes take place:
The text in the title (Main display) changes
to “Standard B” and gives additional advice
that results of standard B are shown in the
adjacent picture. The second light blue line
shows that ASTM 3588 is standard B. To
the right the link
Standard A
is shown,
which enables a changeover again.
Figure 7.33: Examples of the appearance of the basic display with 2 Calculation standards
information for amendment or extension of this setting
See chapter 8